Rapid separation of zirconium using microvolume anion-exchange cartridge for 93Zr determination with isotope dilution ICP-MS

Talanta. 2018 Aug 1:185:98-105. doi: 10.1016/j.talanta.2018.03.026. Epub 2018 Mar 14.

Abstract

Estimating the risks associated with radiation from long-lived fission products (LLFP) in radioactive waste is essential to ensure the long-term safety of potential disposal sites. In this study, the amount of 93Zr, a LLFP, was determined by ICP-MS after separating Zr from a spent nuclear fuel solution using a microvolume anion-exchange cartridge (TEDA cartridge). Zirconium in 9.4 M HCl was stably retained on the TEDA cartridge and readily eluted with 0.75 mL of a mixed solution of 9.4 M HCl and 0.01 M HF. The time taken to complete the Zr separation was 1.2 min. Almost all the other elements initially present in the spent nuclear fuel sample were removed, leading to accurate measurement of all six Zr isotopes (90Zr, 91Zr, 92Zr, 93Zr, 94Zr, and 96Zr). This demonstrated that the TEDA cartridge allowed highly selective separation of Zr regardless of its small bed volume of 0.08 cm3. The concentrations of these isotopes were determined by an isotope-dilution method using a natural Zr standard that has a different isotopic composition from that of the spent nuclear fuel sample. The amount of 93Zr in an initial spent nuclear fuel pellet was 1081 ± 79 ng per mg of 238U. The measured concentrations of all Zr isotopes, as well as the isotopic composition, were consistent with values predicted using a burnup calculation code.

Keywords: (93)Zr; Anion-exchange; Burnup calculation code; Disk cartridge; Isotope dilution ICP-MS; Spent nuclear fuel; TEDA.