Speed Modulation of the HeartWare HVAD to Assess In Vitro Hemocompatibility of Pulsatile and Continuous Flow Regimes in a Rotary Blood Pump

Artif Organs. 2018 Sep;42(9):879-890. doi: 10.1111/aor.13142. Epub 2018 May 3.

Abstract

Although rotary blood pumps (RBPs) sustain life, blood exposure to continuous supra-physiological shear stress induces adverse effects (e.g., thromboembolism); thus, pulsatile flow in RBPs represents a potential solution. The present study introduced pulsatile flow to the HeartWare HVAD using a custom-built controller and compared hemocompatibility biomarkers (i.e., platelet aggregation, concentrations for ADAMTS13, von Willebrand factor (vWf), and free-hemoglobin in plasma (pfHb), red blood cell (RBC) deformability, and RBC-nitric oxide synthase (NOS) activity) between continuous and pulsatile flow in a blood circulation loop over 5 h. The HeartWare HVAD was operated using a custom-built controller, at continuous speed (3282 rev/min) or in a pulsatile mode (mean speed = 3273 rev/min, amplitude = 430 rev/min, frequency = 1 Hz) to generate a blood flow rate of 5.0 L/min, HVAD differential pressure of 90 mm Hg for continuous flow and 92 mm Hg for pulsatile flow, and systolic and diastolic pressures of 121/80 mm Hg. For both flow regimes, the current study found; (i) ADP- and collagen-induced platelet aggregation, and ADAMTS13 concentration significantly decreased after 5 h (P < 0.01; P < 0.05), (ii) ristocetin-induced platelet aggregation significantly increased after 45 min (P < 0.05), (iii) vWf concentration did not significantly differ at any time point, (iv) pfHb significantly increased after 5 h (P < 0.01), (v) RBC deformability improved during the continuous flow regime (P < 0.05) but not during pulsatile flow, and (vi) RBC-NOS activity significantly increased during continuous flow (15 min), and pulsatile flow (5 h; P < 0.05). The current study demonstrated: (i) speed modulation does not improve hemocompatibility of the HeartWare HVAD based on no observable differences being detected for routine biomarkers, and (ii) the time-course for increased RBC-NOS activity observed during continuous flow may have improved RBC deformability.

Keywords: -Erythrocyte; -Hemolysis; -Mechanical circulatory support; -Pulsatility; -Rheology; -Rotary blood pump; -Shear stress; Ventricular assist device.

MeSH terms

  • Adult
  • Erythrocyte Deformability
  • Erythrocytes / physiology*
  • Heart-Assist Devices*
  • Hemodynamics / physiology*
  • Humans
  • Male
  • Models, Cardiovascular
  • Pulsatile Flow / physiology
  • Rheology
  • Stress, Mechanical