Performance of cardiopulmonary resuscitation feedback systems in a long-distance train with distributed traction

Technol Health Care. 2018;26(3):529-535. doi: 10.3233/THC-181241.

Abstract

Background: Out-of-hospital cardiac arrest is common in public locations, including public transportation sites. Feedback devices are increasingly being used to improve chest-compression quality. However, their performance during public transportation has not been studied yet.

Objective: To test two CPR feedback devices representative of the current technologies (accelerometer and electromag- netic-field) in a long-distance train.

Methods: Volunteers applied compressions on a manikin during the train route using both feedback devices. Depth and rate measurements computed by the devices were compared to the gold-standard values.

Results: Sixty-four 4-min records were acquired. The accelerometer-based device provided visual help in all experiments. Median absolute errors in depth and rate were 2.4 mm and 1.3 compressions per minute (cpm) during conventional speed, and 2.5 mm and 1.2 cpm during high speed. The electromagnetic-field-based device never provided CPR feedback; alert messages were shown instead. However, measurements were stored in its internal memory. Absolute errors for depth and rate were 2.6 mm and 0.7 cpm during conventional speed, and 2.6 mm and 0.7 cpm during high speed.

Conclusions: Both devices were accurate despite the accelerations and the electromagnetic interferences induced by the train. However, the electromagnetic-field-based device would require modifications to avoid excessive alerts impeding feedback.

Keywords: CPR feedback; CPR quality; Cardiopulmonary resuscitation (CPR); chest compressions; public access defibrillation (PAD); train.