Biomechanical comparison of glenoid implants with adaptable and fixed backside curvatures in anatomic total shoulder arthroplasty

J Shoulder Elbow Surg. 2018 Sep;27(9):1656-1663. doi: 10.1016/j.jse.2018.02.050. Epub 2018 Apr 27.

Abstract

Background: We evaluated the biomechanical effects and potential advantages of glenoid implants with adaptable backside curvature radii and compared them with standard implants having fixed backside curvatures in anatomic total shoulder arthroplasty (aTSA) for primary glenohumeral osteoarthritis with uniconcave glenoids.

Methods: A glenoid implant with adaptable backside curvatures (Aequalis PerFORM, Tornier SAS, Montbonnot, France) was compared with its previous model having a fixed curvature radius. Virtual aTSAs were performed in 24 patients from preoperative shoulder computed tomography data sets, using both implants in each patient. For all 48 simulated aTSAs, we first measured the glenoid bone reaming depth, subchondral bone quality after reaming, and implant backside surface and then the predicted cement stress, bone-cement interfacial stress, and bone strain at 60° of arm abduction. These biomechanical quantities were tested for differences between adaptable and fixed implants and for correlations between preoperative measurements and postoperative predictions.

Results: Adaptable glenoid implants induced a significant decrease in cement stress (P = .008), bone-cement interfacial stress (P = .045), and bone strain (P = .039), particularly for glenoids with curvature radii larger than 40 mm. However, these biomechanical effects were not significantly correlated with an increase in subchondral glenoid bone quality.

Conclusions: Our study confirms the presumed biomechanical advantages of adaptable glenoid implants, even though the effects were not directly due to the adaptation of the backside curvature radius. Benefits were more pronounced for glenoids with large curvature radii. Our initial biomechanical findings should now be corroborated with large-scale clinical studies.

Keywords: Anatomic total shoulder arthroplasty; aseptic loosening; bone mineral density; bone quality; computed tomography; glenoid implant.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Shoulder / instrumentation*
  • Arthroplasty, Replacement, Shoulder / methods
  • Bone Cements
  • Female
  • Humans
  • Male
  • Middle Aged
  • Osteoarthritis / surgery*
  • Prosthesis Design
  • Scapula / surgery*
  • Shoulder Joint*
  • Shoulder Prosthesis*
  • Tomography, X-Ray Computed

Substances

  • Bone Cements