Automated objective robot-assisted assessment of wrist passive ranges of motion

J Biomech. 2018 May 17:73:223-226. doi: 10.1016/j.jbiomech.2018.03.001. Epub 2018 Mar 13.

Abstract

The measurement of wrist passive ranges of motion (ROMs) can provide insight into improvements and allow for effective monitoring during a rehabilitation program. Compared with conventional methods, this study proposed a new robotic assessment technique for measuring passive ROMs of the wrist. The robotic system has a reconfigurable handle structure that allows for multi-dimensional applications of wrist motions. The assessment reliability of this robotic system was analysed on 11 subjects for measuring wrist extension/flexion and radial/ulnar deviation. Preliminary data demonstrated its potential with intraclass correlation coefficient (ICC2,1) all greater than 0.857 and standard error of measurement (SEM) less than 3.38°. Future work will focus on the standardization of the assessment protocol of this robotic system for assessment purposes, paving the way for its clinical application.

Keywords: Assessment technique; Passive range of motion; Robotic; Wrist.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Female
  • Humans
  • Male
  • Radius / physiology
  • Range of Motion, Articular
  • Reproducibility of Results
  • Robotics*
  • Ulna / physiology
  • Wrist Joint / physiology*
  • Young Adult