Day-and-Night Closed-Loop Insulin Delivery in a Broad Population of Pregnant Women With Type 1 Diabetes: A Randomized Controlled Crossover Trial

Diabetes Care. 2018 Jul;41(7):1391-1399. doi: 10.2337/dc17-2534. Epub 2018 Mar 13.

Abstract

Objective: Despite advances in technology, optimal glucose control remains elusive and neonatal complications remain ubiquitous in type 1 diabetes (T1D) pregnancy. Our aim was to examine the safety, efficacy, and longer-term feasibility of day-and-night closed-loop insulin delivery.

Research design and methods: We recruited 16 pregnant women (mean [SD]: age 32.8 [5.0] years, T1D duration 19.4 [10.2] years, HbA1c 8.0% [1.1], and BMI 26.6 [4.4] kg/m2) to an open-label, randomized, crossover trial. Participants completed 28 days of closed-loop and sensor-augmented pump (SAP) insulin delivery separated by a washout period. Afterward, participants could continue to use the closed-loop system up to 6 weeks postpartum. The primary end point was the proportion of time with glucose levels within the target range (63-140 mg/dL).

Results: The proportion of time with glucose levels within target was comparable during closed-loop and SAP insulin delivery (62.3 vs. 60.1% [95% CI -4.1 to 8.3]; P = 0.47). Mean glucose and time spent hyperglycemic >140 mg/dL also did not differ (131.4 vs. 131.4 mg/dL [P = 0.85] and 36.6 vs. 36.1% [P = 0.86], respectively). During closed-loop, fewer hypoglycemic episodes occurred (median 8 [range 1-17] vs. 12.5 [1-53] over 28 days; P = 0.04) and less time at <63 mg/dL (1.6 vs. 2.7%; P = 0.02). Hypoglycemia <50 mg/dL (0.24 vs. 0.47%; P = 0.03) and low blood glucose index (1.0 vs. 1.4; P = 0.01) were lower. Less nocturnal hypoglycemia (2300-0700 h) during closed-loop therapy (1.1 vs. 2.7%; P = 0.008) and a trend toward higher overnight time in target (67.7 vs. 60.6%; P = 0.06) were found.

Conclusions: Closed-loop insulin delivery was associated with comparable glucose control and significantly less hypoglycemia than SAP therapy. Larger, longer-duration multicenter trials are now indicated to determine clinical efficacy of closed-loop insulin delivery in T1D pregnancy and the impact on neonatal outcomes.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Blood Glucose / metabolism
  • Circadian Rhythm
  • Cross-Over Studies
  • Delivery, Obstetric / statistics & numerical data
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Diabetes Mellitus, Type 1 / epidemiology
  • Female
  • Humans
  • Hyperinsulinism / epidemiology
  • Hypoglycemia / epidemiology
  • Hypoglycemic Agents / administration & dosage*
  • Insulin / administration & dosage*
  • Insulin Infusion Systems*
  • Middle Aged
  • Postpartum Period
  • Pregnancy
  • Pregnancy Outcome / epidemiology
  • Pregnancy in Diabetics / blood
  • Pregnancy in Diabetics / drug therapy*
  • Time Factors
  • Treatment Outcome
  • Young Adult

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin

Associated data

  • ISRCTN/ISRCTN83316328