A Sol-gel Integrated Dual-readout Microarray Platform for Quantification and Identification of Prostate-specific Antigen

Anal Sci. 2018;34(3):317-321. doi: 10.2116/analsci.34.317.

Abstract

Here, we report a sol-gel integrated affinity microarray for on-chip matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF-MS) that enables capture and identification of prostate?specific antigen (PSA) in samples. An anti-PSA antibody (H117) was mixed with a sol?gel, and the mixture was spotted onto a porous silicon (pSi) surface without additional surface modifications. The antibody easily penetrates the sol-gel macropore fluidic network structure, making possible high affinities. To assess the capture affinity of the platform, we performed a direct assay using fluorescein isothiocyanate-labeled PSA. Pure PSA was subjected to on-chip MALDI-TOF-MS analysis, yielding three clear mass peptide peaks (m/z = 1272, 1407, and 1872). The sol-gel microarray platform enables dual readout of PSA both fluorometric and MALDI-TOF MS analysis in biological samples. Here we report a useful method for a means for discovery of biomarkers in complex body fluids.

Keywords: MALDI-TOF mass spectrometry; PSA; Sol-gel microarray; biomarker discovery; macro-structure silicon; sensitive diagnosis.

MeSH terms

  • Gels
  • Microarray Analysis / instrumentation*
  • Prostate-Specific Antigen / analysis*
  • Prostate-Specific Antigen / metabolism
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization
  • Trypsin / metabolism

Substances

  • Gels
  • Trypsin
  • Prostate-Specific Antigen