Continuous stacking computational approach based automated microscope slide scanner

Rev Sci Instrum. 2018 Feb;89(2):023701. doi: 10.1063/1.5022549.

Abstract

Cost-effective and automated acquisition of whole slide images is a bottleneck for wide-scale deployment of digital pathology. In this article, a computation augmented approach for the development of an automated microscope slide scanner is presented. The realization of a prototype device built using inexpensive off-the-shelf optical components and motors is detailed. The applicability of the developed prototype to clinical diagnostic testing is demonstrated by generating good quality digital images of malaria-infected blood smears. Further, the acquired slide images have been processed to identify and count the number of malaria-infected red blood cells and thereby perform quantitative parasitemia level estimation. The presented prototype would enable cost-effective deployment of slide-based cyto-diagnostic testing in endemic areas.

MeSH terms

  • Automation
  • Equipment Design
  • Image Processing, Computer-Assisted
  • Malaria / diagnostic imaging
  • Microscopy / instrumentation*