Enhancement of measurement accuracy of X-ray PIV in comparison with the micro-PIV technique

J Synchrotron Radiat. 2018 Mar 1;25(Pt 2):552-559. doi: 10.1107/S1600577517017398. Epub 2018 Feb 13.

Abstract

The X-ray PIV (particle image velocimetry) technique has been used as a non-invasive measurement modality to investigate the haemodynamic features of blood flow. However, the extraction of two-dimensional velocity field data from the three-dimensional volumetric information contained in X-ray images is technically unclear. In this study, a new two-dimensional velocity field extraction technique is proposed to overcome technological limitations. To resolve the problem of finding a correction coefficient, the velocity field information obtained by X-ray PIV and micro-PIV techniques for disturbed flow in a concentric stenosis with 50% severity was quantitatively compared. Micro-PIV experiments were conducted for single-plane and summation images, which provide similar positional information of particles as X-ray images. The correction coefficient was obtained by establishing the relationship between velocity data obtained from summation images (VS) and centre-plane images (VC). The velocity differences between VS and VC along the vertical and horizontal directions were quantitatively analysed as a function of the geometric angle of the test model for applying the present two-dimensional velocity field extraction technique to a conduit of arbitrary geometry. Finally, the two-dimensional velocity field information at arbitrary positions could be successfully extracted from X-ray images by using the correction coefficient and several velocity parameters derived from VS.

Keywords: X-ray PIV; biofluid flow; correction coefficient; micro-PIV; stenosis.