Blood glucose monitoring during aerobic and anaerobic physical exercise using a new artificial pancreas system

Endocrinol Diabetes Nutr (Engl Ed). 2018 Jun-Jul;65(6):342-347. doi: 10.1016/j.endinu.2017.12.012. Epub 2018 Feb 23.
[Article in English, Spanish]

Abstract

Aim: To assess an artificial pancreas system during aerobic (AeE) and anaerobic exercise (AnE).

Methods: A pilot clinical trial on five subjects with type 1 diabetes (4 males) aged 37±10.9 years, diabetes diagnosed 21.2±12.2 years before, insulin pump users, and with a mean HbA1c level of 7.8±0.5%. Every subject did three AeE and three AnE sessions. Blood glucose levels were monitored by the artificial pancreas system during exercise and up to four hours later. Before the start of exercise, 23g of carbohydrates were administered orally.

Results: The mean glucose level was 124.0±25.1mg/dL in the AeE studies and 152.1±34.1mg/dL in the AnE studies. Percent times in the different glucose ranges of 70-180, >180 and <70mg/dL were 89.8±18.6% and 75.9±27.6%; 7.7±18.4% and 23.2±28.0%; and 2.5±6.3% and 1.0±3.6% during the AeE and AnE sessions, respectively. Only six rescues with carbohydrates (15g) were required during the studies (4 in AeE and 2 in AnE). Total insulin dose during the five hours of the study was 3.1±1.0IU in the AeE studies and 3.5±1.3IU in the AnE studies.

Conclusions: Blood glucose response to AeE and AnE exercise is different. The evaluated artificial pancreas system appeared to achieve effective and safe blood glucose control during exercise and up to four hours later. However, new control strategies that minimize patient intervention should be designed.

Keywords: Artificial pancreas; Diabetes tipo 1; Ejercicio; Exercise; Páncreas artificial; Type 1 diabetes.