Lesion detection and quantification performance of the Tachyon-I time-of-flight PET scanner: phantom and human studies

Phys Med Biol. 2018 Mar 16;63(6):065010. doi: 10.1088/1361-6560/aab0f3.

Abstract

The first generation Tachyon PET (Tachyon-I) is a demonstration single-ring PET scanner that reaches a coincidence timing resolution of 314 ps using LSO scintillator crystals coupled to conventional photomultiplier tubes. The objective of this study was to quantify the improvement in both lesion detection and quantification performance resulting from the improved time-of-flight (TOF) capability of the Tachyon-I scanner. We developed a quantitative TOF image reconstruction method for the Tachyon-I and evaluated its TOF gain for lesion detection and quantification. Scans of either a standard NEMA torso phantom or healthy volunteers were used as the normal background data. Separately scanned point source and sphere data were superimposed onto the phantom or human data after accounting for the object attenuation. We used the bootstrap method to generate multiple independent noisy datasets with and without a lesion present. The signal-to-noise ratio (SNR) of a channelized hotelling observer (CHO) was calculated for each lesion size and location combination to evaluate the lesion detection performance. The bias versus standard deviation trade-off of each lesion uptake was also calculated to evaluate the quantification performance. The resulting CHO-SNR measurements showed improved performance in lesion detection with better timing resolution. The detection performance was also dependent on the lesion size and location, in addition to the background object size and shape. The results of bias versus noise trade-off showed that the noise (standard deviation) reduction ratio was about 1.1-1.3 over the TOF 500 ps and 1.5-1.9 over the non-TOF modes, similar to the SNR gains for lesion detection. In conclusion, this Tachyon-I PET study demonstrated the benefit of improved time-of-flight capability on lesion detection and ROI quantification for both phantom and human subjects.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Abdominal Neoplasms / diagnosis*
  • Abdominal Neoplasms / diagnostic imaging
  • Heart Neoplasms / diagnosis*
  • Heart Neoplasms / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted
  • Lutetium / chemistry
  • Phantoms, Imaging*
  • Positron-Emission Tomography / instrumentation*
  • Positron-Emission Tomography / methods*
  • Radiography, Thoracic
  • Signal-To-Noise Ratio
  • Silicates / chemistry
  • Time Factors

Substances

  • Silicates
  • lutetium orthosilicate
  • Lutetium