Preclinical evaluation of a semi-automated and rapid commercial electrophoresis assay for von Willebrand factor multimers

J Clin Lab Anal. 2018 Jul;32(6):e22416. doi: 10.1002/jcla.22416. Epub 2018 Feb 17.

Abstract

Background: The von Willebrand factor (VWF) multimer test is required to correctly subtype qualitative type 2 von Willebrand disease (VWD). The current VWF multimer assays are difficult, nonstandardized, and time-consuming. The purpose of this study was to evaluate the clinical utility of the commercial VWF multimer kit by Sebia (Lisses, France), an electrophoresis technique yielding same-day results.

Methods: Ten healthy volunteer plasma samples, in-house reference plasma (IRP) and commercial normal plasma (CNP) samples, 10 plasma samples from patients with a known VWD type, 1 hemophilia A plasma sample, and 7 external quality assurance (EQA) samples were analyzed using the commercial VWF multimer kit. Additional coagulation testing included measurements of VWF antigen (VWF:Ag), VWF activity (VWF:Ac), and FVIII activity (FVIII:C).

Results: The CNP results revealed a relative loss of the highest molecular weight multimers; therefore, IRP was preferred as the reference sample. The interpretations of 10 patients with a known VWD type could be successfully reproduced and agreed with previous VWF multimer results. In all EQA surveys, the multimer results and final VWD diagnosis agreed with expert opinion.

Conclusions: The VWF multimer assay by Sebia is easy to perform and can be successfully implemented in any clinical laboratory for second-stage evaluation of VWD. The resolution power of multimer distribution is adequate to correctly classify VWD types 1, 2A, 2B, and 3.

Keywords: electrophoresis; subtyping; von Willebrand disease; von Willebrand factor; von Willebrand factor multimer.