Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes

Pediatr Cardiol. 2018 Mar;39(3):627-632. doi: 10.1007/s00246-018-1828-5. Epub 2018 Feb 17.

Abstract

Transcatheter closure of arterial duct (AD) remains challenging in low-weight patients and using Amplatzer Duct Occluder devices is still considered off-label in infants < 6 kg. This study aimed to report a large, single-center experience of percutaneous AD closure in infants < 6 kg as well as to compare the most frequently used devices, Amplatzer Duct Occluder type I (ADO I) and Amplatzer Duct Occluder type II Additional Sizes devices (ADO II-AS) (St. Jude Medical Corp, St. Paul, MN, USA). From March 2000 to March 2017, among the 762 patients submitted to percutaneous closure of AD at our Institution, 33 were infants < 6 kg (age 4.8 ± 2.1 months; weight 5.0 ± 0.9 kg). Fourteen patients (45%) underwent ADO I (Group I) and 19 patients ADO II-AS (Group II) device implantation. AD diameter was 2.6 ± 0.8 (range 1.5-4.0) mm resulting in QP/QS of 2.6 ± 0.0.9 (range 1.1-4.5). Successful device deployment was achieved in all patients without procedural morbidity or mortality. Procedural and fluoroscopy times were not significantly different between the groups. However, total X-ray absorbed dose was significantly higher in Group I (121 ± 69 vs 29 ± 16 mGy/cm2, p < 0.01). Immediate, 24 h, and mid-term (46 ± 37 months, median 39) complete occlusion were recorded in 72.7, 90.9, and 97% of patients, respectively without significant difference between the groups. In conclusion, transcatheter closure of AD with Amplatzer Duct Occluder devices is feasible, safe, and effective also in infants less than 6 kg, without significant difference between the most commonly used devices, namely, ADO I and ADO II-AS.

Keywords: Arterial duct; Device; Low-weight infant.

Publication types

  • Comparative Study

MeSH terms

  • Cardiac Catheterization / instrumentation*
  • Ductus Arteriosus, Patent / classification
  • Ductus Arteriosus, Patent / diagnostic imaging
  • Ductus Arteriosus, Patent / surgery*
  • Female
  • Fluoroscopy
  • Humans
  • Infant
  • Infant, Low Birth Weight
  • Male
  • Prosthesis Design
  • Radiography
  • Septal Occluder Device*
  • Treatment Outcome