Treatment of Aortic Stenosis With a Self-Expanding, Resheathable Transcatheter Valve: One-Year Results of the International Multicenter Portico Transcatheter Aortic Valve Implantation System Study

Circ Cardiovasc Interv. 2018 Feb;11(2):e005206. doi: 10.1161/CIRCINTERVENTIONS.117.005206.

Abstract

Background: The aim of the Portico TAVI (transcatheter aortic valve implantation) system study was to evaluate outcomes ≤1 year after implantation of a novel resheathable, self-expanding TAVI system in a multicenter patient population with severe aortic stenosis (AS).

Methods and results: High-risk patients (n=222) with symptomatic severe AS (mean age, 83.0±4.6 years; 74.3% women) were enrolled across 12 centers in Europe and Australia. The study's primary end point was all-cause mortality at 30 days. A total of 209 patients who received the Portico TAVI system were available for follow-up after the 30-day visit. Data collection included hemodynamic assessment by echocardiography with core laboratory evaluation and assessment of functional status. Valve Academic Research Consortium-defined adverse events were adjudicated by an independent Clinical Events Committee. TAVI using the Portico valve led to a significant and persistent improvement in aortic valve function at 1 year. More than mild paravalvular leak was present in 5.7% and 7.5% of patients at 30 days and 1 year, respectively. Kaplan-Meier estimates at 30 days and 1 year were 3.6% and 13.8% for all-cause mortality, 3.6% and 9.6% for cardiovascular mortality, and 3.2% and 5.8% for major (disabling) stroke. After 30 days and ≤1 year of follow-up, adverse events included stage 3 acute kidney injury (n=3), major vascular complications (n=5), and life-threatening/disabling bleeding (n=3). Overall permanent pacemaker rate was 14.7%. At 1 year, 74.8% improved ≥1 New York Heart Association class compared with baseline (P<0.0001).

Conclusions: The Portico TAVI system is safe and effective at 1 year, yielding low mortality and stroke rates in high-risk patients with severe AS.

Clinical trial registration: URL: https://www.clinicaltrials.gov. Unique identifier: NCT01493284.

Keywords: aortic valve stenosis; follow-up studies; humans; self-expanding valve; transcatheter.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Australia / epidemiology
  • Cardiac Catheters*
  • Europe / epidemiology
  • Female
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Postoperative Complications / epidemiology
  • Prospective Studies
  • Prosthesis Design
  • Recovery of Function
  • Severity of Illness Index
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01493284