Robust Ionic Current Sensor for Bacterial Cell Size Detection

ACS Sens. 2018 Mar 23;3(3):574-579. doi: 10.1021/acssensors.8b00045. Epub 2018 Mar 5.

Abstract

Ionic current sensing methods are useful tools for detecting sub- to several-micron scale particles such as bacteria. However, conventional commercially available ionic current sensing devices are not suitable for on-site measurement use because of inherent limitations on their robustness. Here, we proposed a portable robust ionic current sensor (Robust-ICS) using a bridge circuit that offers a high signal-to-noise (S/N) ratio by suppressing background current. Because the Robust-ICS can tolerate increased noise in current sensing, a simple, lightweight electromagnetic shield can be used and measurements under large electromagnetic noise conditions can be made. The weight of the device was lowered below 4 kg and outdoor particle detection measurements were completed successfully. Accuracy of size detection of Staphylococcus aureus ( S. aureus) was equivalent to that obtained by SEM imaging.

Keywords: bacteria; bridge circuit; ionic current sensing; micropore; robust sensor.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electromagnetic Phenomena*
  • Particle Size
  • Staphylococcus aureus / cytology*
  • Surface Properties