Design of a new artificial breathing system for simulating the human respiratory activities

J Med Eng Technol. 2018 Jan;42(1):52-58. doi: 10.1080/03091902.2018.1430185. Epub 2018 Feb 7.

Abstract

The purpose of this work is the conception and implementation of an artificial active respiratory system that allows the simulation of human respiratory activities. The system consists of two modules, mechanical and electronical. The first one represents a cylindrical lung adjustable in resistance and compliance. This lung is located inside a transparent thoracic box, connected to a piston that generates variable respiratory efforts. The parameters of the system, which are pressure, flow and volume, are measured by the second module. A computer application was developed to control the whole system, and enables the display of the parameters. A series of tests were made to evaluate the respiratory efforts, resistances and compliances. The results were compared to the bibliographical studies, allowing the validation of the proposed system.

Keywords: Respiratory system simulation; active test lung; lung compliance; lung resistance; respiratory efforts.

MeSH terms

  • Artificial Organs*
  • Computer Simulation
  • Equipment Design
  • Humans
  • Lung / physiology*
  • Models, Biological*
  • Pressure
  • Respiration, Artificial / instrumentation*