Development and validation of RP-HPLC method for estimation of eplerenone in spiked human plasma

J Pharm Anal. 2012 Oct;2(5):390-393. doi: 10.1016/j.jpha.2012.04.006. Epub 2012 Apr 24.

Abstract

A rapid and simple high performance liquid chromatography (HPLC) method with a UV detection (241 nm) was developed and validated for estimation of eplerenone from spiked human plasma. The analyte and the internal standard (valdecoxib) were extracted with a mixture of dichloromethane and diethyl ether. The chromatographic separation was performed on a HiQSil C-18HS column (250 mm×4.6 mm, 5 μm) with a mobile phase consisting of acetonitrile:water (50:50, v/v) at flow rate of 1 mL/min. The calibration curve was linear in the range 100-3200 ng/mL and the heteroscedasticity was minimized by using weighted least squares regression with weighting factor 1/X.

Keywords: Eplerenone; HPLC–UV; Liquid–liquid extraction; Weighted regression.