Mid-infrared upconversion based hyperspectral imaging

Opt Express. 2018 Feb 5;26(3):2203-2211. doi: 10.1364/OE.26.002203.

Abstract

Mid-infrared hyperspectral imaging has in the past decade emerged as a promising tool for medical diagnostics. In this work, nonlinear frequency upconversion based hyperspectral imaging in the 6 to 8 µm spectral range is presented for the first time, using both broadband globar and narrowband quantum cascade laser illumination. AgGaS2 is used as the nonlinear medium for sum frequency generation using a 1064 nm mixing laser. Angular scanning of the nonlinear crystal provides broad spectral coverage at every spatial position in the image. This study demonstrates the retrieval of series of monochromatic images acquired by a silicon based CCD camera, using both broadband and narrowband illumination and a comparison is made between the two illumination sources for hyperspectral imaging.