Subaxillary approach to cardiac electronic device implantation using a single surgical incision: a single-centre experience

Interact Cardiovasc Thorac Surg. 2018 Jun 1;26(6):912-918. doi: 10.1093/icvts/ivy002.

Abstract

Objectives: Pacemakers and implantable cardioverter defibrillators have become more common in younger individuals, owing to broader indications and technological advances. Our goal was to report our long-term experience of implanting cardiac devices in young adults via a subaxillary approach with a hidden single incision, to minimize the aesthetic impact of the device.

Methods: From 2003 to 2012, 23 patients underwent cardiac device implantation via a subaxillary approach while they were under general anaesthesia. The axillary vein was punctured under contrast-medium radiological guidance. The devices were positioned into the pocket under the pectoralis major muscle and over the fascia that envelopes the pectoralis minor muscle, through a lateral surgical access along the midaxillary line.

Results: Twenty-two single- and dual-chamber implantable cardioverter defibrillator and 1 dual-chamber pacemaker were implanted (mean age 32.2 ± 13.4 years). All procedures were successfully performed without complications; overall hospitalization after implantation was 3.3 days. During mean a follow-up of 65 ± 18 months, no deaths, infections or device migrations occurred. Two ventricular lead dislodgements requiring repositioning occurred within 1 month, and 1 defibrillation lead fracture occurred 5 years after implantation. All patients were extremely satisfied with the aesthetic result. A mild increase in the pacing threshold (0.88 ± 0.6 V at 0.4/0.5 ms at implantation vs 1.00 ± 0.3 V at 0.4/0.5 ms at 2-year follow-up, P = 0.063) and a statistically significant decrease in the pacing impedance (637 ± 161 ohm at implantation vs 499 ± 81 ohm at 2-year follow-up, P = 0.001) were observed, without any consequences.

Conclusions: The subaxillary approach is a safe, feasible technique that requires a simple surgical procedure. The implantable cardioverter defibrillator functions properly at implantation and at long-term follow-up.

MeSH terms

  • Adolescent
  • Adult
  • Anesthesia, General
  • Arrhythmias, Cardiac / therapy*
  • Axilla
  • Axillary Vein / surgery*
  • Defibrillators, Implantable*
  • Female
  • Hospitalization
  • Humans
  • Male
  • Middle Aged
  • Pectoralis Muscles
  • Prosthesis Implantation / methods*
  • Surgical Wound*
  • Young Adult