A Universal Standard for the Validation of Blood Pressure Measuring Devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement

Hypertension. 2018 Mar;71(3):368-374. doi: 10.1161/HYPERTENSIONAHA.117.10237. Epub 2018 Jan 31.

Abstract

In the past 30 years, several organizations, such as the US Association for the Advancement of Medical Instrumentation (AAMI), the British Hypertension Society, the European Society of Hypertension (ESH) Working Group on Blood Pressure (BP) Monitoring, and the International Organization for Standardization (ISO), have developed protocols for clinical validation of BP measuring devices. However, it is recognized that science, as well as patients, consumers, and manufacturers, would be best served if all BP measuring devices were assessed for accuracy according to an agreed single validation protocol that had global acceptance. Therefore, an international initiative was taken by the AAMI, ESH, and ISO experts who agreed to develop a universal standard for device validation. This statement presents the key aspects of a validation procedure, which were agreed by the AAMI, ESH, and ISO representatives as the basis for a single universal validation protocol. As soon as the AAMI/ESH/ISO standard is fully developed, this will be regarded as the single universal standard and will replace all other previous standards/protocols.

Keywords: accuracy; blood pressure measurement; device; monitor; protocol; standard; validation.

Publication types

  • Review

MeSH terms

  • Blood Pressure Determination / standards*
  • Blood Pressure Monitors / standards*
  • Consensus*
  • Europe
  • Humans
  • International Cooperation
  • Practice Guidelines as Topic / standards*
  • Reference Standards
  • Reproducibility of Results
  • Societies, Medical / standards