Evaluation of a novel thoracic entry device versus needle decompression in a tension pneumothorax swine model

Am J Surg. 2018 May;215(5):832-835. doi: 10.1016/j.amjsurg.2017.12.014. Epub 2018 Jan 5.

Abstract

Introduction: Tension pneumothorax (tPTX) remains a major cause of preventable death in trauma. Needle decompression (ND) has up to a 60% failure rate.

Methods: Post-mortem swine used. Interventions were randomized to 14G-needle decompression (ND, n = 25), bladed trocar with 36Fr cannula (BTW, n = 16), bladed trocar alone (BTWO, n = 16) and surgical thoracostomy (ST = 11). Simulated tPTX was created to a pressure(p) of 20 mmHg.

Results: Success (p < 5 mmHg by 120 s) was seen in 41 of 68 (60%) interventions. BTW and BTWO were consistently more successful than ND with success rates of 88% versus 48% in ND (p < .001). In successful deployments, ND was slower to reach p < 5 mmHg, average of 82s versus 26s and 28s for BTW and BTWO respectively (p < .001). Time to implement procedure was faster for ND with an average of 3.6s versus 16.9s and 15.3s in the BTW and BTWO (p < .001). Final pressure was significantly less in BTW and BTWO at 1.7 mmHg versus 7 mmHg in ND animals (p < .001).

Conclusion: Bladed trocars can safely and effectively tPTX with a significantly higher success rates than needle decompression.

Keywords: Bladed trocar thoracostomy; Needle decompression; Reactor; Tension pneumothorax.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Decompression, Surgical / instrumentation*
  • Disease Models, Animal
  • Equipment Design
  • Needles
  • Pneumothorax / surgery*
  • Surgical Instruments*
  • Swine
  • Thoracostomy / methods