Motion Artifact Quantification and Sensor Fusion for Unobtrusive Health Monitoring

Sensors (Basel). 2017 Dec 25;18(1):38. doi: 10.3390/s18010038.

Abstract

Sensors integrated into objects of everyday life potentially allow unobtrusive health monitoring at home. However, since the coupling of sensors and subject is not as well-defined as compared to a clinical setting, the signal quality is much more variable and can be disturbed significantly by motion artifacts. One way of tackling this challenge is the combined evaluation of multiple channels via sensor fusion. For robust and accurate sensor fusion, analyzing the influence of motion on different modalities is crucial. In this work, a multimodal sensor setup integrated into an armchair is presented that combines capacitively coupled electrocardiography, reflective photoplethysmography, two high-frequency impedance sensors and two types of ballistocardiography sensors. To quantify motion artifacts, a motion protocol performed by healthy volunteers is recorded with a motion capture system, and reference sensors perform cardiorespiratory monitoring. The shape-based signal-to-noise ratio SNR S is introduced and used to quantify the effect on motion on different sensing modalities. Based on this analysis, an optimal combination of sensors and fusion methodology is developed and evaluated. Using the proposed approach, beat-to-beat heart-rate is estimated with a coverage of 99.5% and a mean absolute error of 7.9 ms on 425 min of data from seven volunteers in a proof-of-concept measurement scenario.

Keywords: ambient assisted living; biosignals; heart rate; medical signal processing; motion artifacts; motion capture; sensor fusion; unobtrusive sensing.

MeSH terms

  • Algorithms
  • Artifacts
  • Ballistocardiography
  • Heart Rate
  • Humans
  • Motion*
  • Photoplethysmography
  • Signal Processing, Computer-Assisted
  • Signal-To-Noise Ratio