Efficacy of a one-catheter concept for transradial coronary angiography

PLoS One. 2018 Jan 2;13(1):e0189899. doi: 10.1371/journal.pone.0189899. eCollection 2018.

Abstract

Introduction: Transradial coronary angiography (TRC) can be performed with a one-catheter approach for the right and left coronary ostium (R/LCO). We investigated the performance of a special diagnostic catheter widely used for the one-catheter-approach, the Tiger (Tiger II, TerumoTM).

Methods: In a dual center registry we analyzed 1412 TRC-procedures exclusively performed by experienced TRC-operators. We compared the performance of the Tiger with Judkins catheters by retrospectively judging ostial catheter stability during contrast injection, and by measuring contrast use, fluoroscopy time (FT) and complication rate.

Results: Poor or failed ostial engagement was found in 40.5% in the Tiger group, compared to 46.6% with the use of Judkins catheters (p<0.183). Ostial instability of the Tiger was found more often during engagement of the LCO than the RCO (34.4% vs. 10.8%, p<0.001), whereas it was similar in the LCO and RCO for Judkins catheters (27.4% vs. 26.7%, p = 0.840). TRC-procedures performed with Tiger catheters were associated with less contrast volume (63.48 ± 29.83mL vs. 82.51 ± 56.58mL, p<0.004) and shorter FT than with Judkins catheters. (198.27 ± 194.8sec vs. 326.85 ± 329.70sec). Forearm hematomas occurred less often with the Tiger (1.2% vs. 6.6%, p< 0.02).

Conclusion: The Tiger employed as a single catheter in TRC is an effective tool to achieve lower contrast volume and fluoroscopy time at a low complication rate. Unstable engagement affects predominantly the left coronary artery, but its overall frequency is similar for both, the Tiger and Judkins catheters.

MeSH terms

  • Aged
  • Cardiac Catheters*
  • Coronary Angiography / methods*
  • Female
  • Humans
  • Male
  • Middle Aged

Grants and funding

The authors received no specific funding for this work.