Note: Micro-channel array crucible for isotope-resolved laser spectroscopy of high-temperature atomic beams

Rev Sci Instrum. 2017 Dec;88(12):126101. doi: 10.1063/1.5006457.

Abstract

We present a novel compact design for a multichannel atomic oven which generates collimated beams of refractory atoms for fieldable laser spectroscopy. Using this resistively heated crucible, we demonstrate spectroscopy of an erbium sample at 1300 °C with improved isotopic resolution with respect to a single-channel design. In addition, our oven has a high thermal efficiency. By minimizing the surface area of the crucible, we achieve 2000 °C at 140 W of applied electrical power. As a result, the design does not require any active cooling and is compact enough to allow for its incorporation into fieldable instruments.