[Evaluation of Image Quality of Readout Segmented EPI with Readout Partial Fourier Technique]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2017;73(12):1244-1251. doi: 10.6009/jjrt.2017_JSRT_73.12.1244.
[Article in Japanese]

Abstract

Readout segmented EPI (readout segmentation of long variable echo-trains: RESOLVE) segmented k-space in the readout direction. By using the partial Fourier method in the readout direction, the imaging time was shortened. However, the influence on image quality due to insufficient data sampling is concerned. The setting of the partial Fourier method in the readout direction in each segment was changed. Then, we examined signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and distortion ratio for changes in image quality due to differences in data sampling. As the number of sampling segments decreased, SNR and CNR showed a low value. In addition, the distortion ratio did not change. The image quality of minimum sampling segments is greatly different from full data sampling, and caution is required when using it.

Keywords: projection onto convex sets (POCS); readout partial Fourier; readout segmented EPI (RESOLVE); readout segments.

Publication types

  • Evaluation Study

MeSH terms

  • Echo-Planar Imaging / methods*
  • Fourier Analysis*
  • Signal-To-Noise Ratio