Abstract

Background and objectives: Interventional endoscopic ultrasound (EUS) is a promising novel approach for intravascular interventions. The aim of this study was to assess the feasibility and safety of a EUS-guided intrahepatic portosystemic shunt (EGIPS) with portal pressure gradient measurement in a live porcine model.

Methods: The left hepatic vein (LHV) or the inferior vena cava (IVC) was punctured with a needle that advanced into the portal vein (PV). A guidewire was then inserted into the PV, and a needle knife was used to create an intrahepatic fistula between LHV and PV. Portal pressure was recorded. The fistula was dilated with a balloon and a biliary metal stent was deployed between LHV and PV under sonographic and fluoroscopic observation. A portocavography validated the patency of the stent. Necropsies were realized after euthanasia.

Results: Portosystemic stenting was achieved in 19/21 pigs. Final portocavography confirmed stent patency between PV and LHV or IVC in 17 pigs (efficacy of 81%): Four stents were dysfunctional as two were thrombosed and two were poor positioned. Portal pressure was documented before and after shunting in 20/21 pigs. Necropsies revealed that 19/21 procedures were transesophageal and two were transgastric. Hemoperitoneum and pneumothorax were found in one pig and hemothorax was found in two pigs. Morbidity was 14.2% (3/21 animals).

Conclusion: EGIPS was feasible in 91% of cases, functional in 81%, with 14.2% per procedure morbidity. EGIPS still needs to be assessed in portal hypertension pig models with longer follow-up before being considered as an alternative when the transjugular intrahepatic portosystemic shunt fails.