Real-time detection system for tumor localization during minimally invasive surgery for gastric and colon cancer removal: In vivo feasibility study in a swine model

J Surg Oncol. 2018 Mar;117(4):699-706. doi: 10.1002/jso.24922. Epub 2017 Nov 28.

Abstract

Background and objectives: During minimally invasive surgery (MIS), it is impossible to directly detect marked clips around tumors via palpation. Therefore, we developed a novel method and device using Radio Frequency IDentification (RFID) technology to detect the position of clips during minimally invasive gastrectomy or colectomy.

Methods: The feasibility of the RFID-based detection system was evaluated in an animal experiment consisting of seven swine. The primary outcome was to successfully detect the location of RFID clips in the stomach and colon. The secondary outcome measures were to detect time (time during the intracorporeal detection of the RFID clip), and accuracy (distance between the RFID clip and the detected site).

Results: A total of 25 detection attempts (14 in the stomach and 11 in the colon) using the RFID antenna had a 100% success rate. The median detection time was 32.5 s (range, 15-119 s) for the stomach and 28.0 s (range, 8-87 s) for the colon. The median detection distance was 6.5 mm (range, 4-18 mm) for the stomach and 6.0 mm (range, 3-13 mm) for the colon.

Conclusions: We demonstrated favorable results for a RFID system that detects the position of gastric and colon tumors in real-time during MIS.

Keywords: early colon cancer; early gastric cancer; intraoperative tumor localization; minimally invasive surgery; radio frequency identification detector.

MeSH terms

  • Algorithms
  • Animals
  • Colonic Neoplasms / surgery*
  • Computer Systems
  • Disease Models, Animal*
  • Feasibility Studies
  • Minimally Invasive Surgical Procedures / instrumentation
  • Minimally Invasive Surgical Procedures / methods*
  • Radio Frequency Identification Device / methods*
  • Stomach Neoplasms / surgery*
  • Swine
  • Treatment Outcome