Hydrogel electrodes are commonly used for functional and other electrical stimulation applications since the hydrogel layer has been shown to considerably reduce the perception of stimulation compared to dry electrodes. However, these hydrogel electrodes must be changed regularly as they dry out or become contaminated with skin cells and sweat products, thus losing their adhesiveness and resistive properties. Dry electrodes are longer lasting but are more uncomfortable due to unequal current distribution (current hogging). We hypothesise that if current through a dry electrode is equally shared amongst an array of small sub-electrodes, current hogging and thus the sensitivity perceived due to stimulation will be reduced. We constructed an 8 × 8 array of millimetre sized dry electrodes that could either be activated as individual current sources, or together as one large source. A study was performed with 13 participants to investigate the differences in sensation between the two modes of operation. The results showed that 12 out of 13 participants found the new (distributed-constant-current) approach allowed higher stimulation for the same sensation. The differences in sensation between single and multiple sources became larger with higher intensity levels.
Keywords: Array stimulation; Dry electrodes; Electrical stimulation.