Positioning of head and neck patients for proton therapy using proton range probes: a proof of concept study

Phys Med Biol. 2017 Dec 29;63(1):015025. doi: 10.1088/1361-6560/aa9cff.

Abstract

To exploit the full potential of proton therapy, accurate and on-line methods to verify the patient positioning and the proton range during the treatment are desirable. Here we propose and validate an innovative technique for determining patient misalignment uncertainties through the use of a small number of low dose, carefully selected proton pencil beams ('range probes') (RP) with sufficient energy that their residual Bragg peak (BP) position and shape can be measured on exit. Since any change of the patient orientation in relation to these beams will result in changes of the density heterogeneities through which they pass, our hypothesis is that patient misalignments can be deduced from measured changes in Bragg curve (BC) shape and range. As such, a simple and robust methodology has been developed that estimates average proton range and range dilution of the detected residual BC, in order to locate range probe positions with optimal prediction power for detecting misalignments. The validation of this RP based approach has been split into two phases. First we retrospectively investigate its potential to detect translational patient misalignments under real clinical conditions. Second, we test it for determining rotational errors of an anthropomorphic phantom that was systematically rotated using an in-house developed high precision motion stage. Simulations of RPs in these two scenarios show that this approach could potentially predict translational errors to lower than1.5 mm and rotational errors to smaller than 1° using only three or five RPs positions respectively.

MeSH terms

  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Organs at Risk / radiation effects
  • Patient Positioning*
  • Phantoms, Imaging*
  • Proof of Concept Study*
  • Proton Therapy / methods*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy Setup Errors / prevention & control
  • Radiotherapy, Intensity-Modulated / methods
  • Retrospective Studies