Predictive hyperglycemia and hypoglycemia minimization: In-home double-blind randomized controlled evaluation in children and young adolescents

Pediatr Diabetes. 2018 May;19(3):420-428. doi: 10.1111/pedi.12603. Epub 2017 Nov 20.

Abstract

Objective: The primary objective of this trial was to evaluate the feasibility, safety, and efficacy of a predictive hyperglycemia and hypoglycemia minimization (PHHM) system vs predictive low glucose suspension (PLGS) alone in optimizing overnight glucose control in children 6 to 14 years old.

Research design and methods: Twenty-eight participants 6 to 14 years old with T1D duration ≥1 year with daily insulin therapy ≥12 months and on insulin pump therapy for ≥6 months were randomized per night into PHHM mode or PLGS-only mode for 42 nights. The primary outcome was percentage of time in sensor-measured range 70 to 180 mg/dL in the overnight period.

Results: The addition of automated insulin delivery with PHHM increased time in target range (70-180 mg/dL) from 66 ± 11% during PLGS nights to 76 ± 9% during PHHM nights (P<.001), without increasing hypoglycemia as measured by time below various thresholds. Average morning blood glucose improved from 176 ± 28 mg/dL following PLGS nights to 154 ± 19 mg/dL following PHHM nights (P<.001).

Conclusions: The PHHM system was effective in optimizing overnight glycemic control, significantly increasing time in range, lowering mean glucose, and decreasing glycemic variability compared to PLGS alone in children 6 to 14 years old.

Keywords: automated insulin delivery; continuous glucose monitoring; type 1 diabetes.

Publication types

  • Randomized Controlled Trial
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Blood Glucose
  • Child
  • Clinical Alarms
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 1 / drug therapy
  • Double-Blind Method
  • Feasibility Studies
  • Female
  • Humans
  • Hyperglycemia / prevention & control*
  • Hypoglycemia / prevention & control*
  • Insulin Infusion Systems*
  • Male
  • Monitoring, Ambulatory / instrumentation*

Substances

  • Blood Glucose