Is vacuum ultraviolet detector a concentration or a mass dependent detector?

J Chromatogr A. 2017 Dec 29:1530:171-175. doi: 10.1016/j.chroma.2017.11.028. Epub 2017 Nov 17.

Abstract

The vacuum ultraviolet detector (VUV) is a very effective tool for chromatogram deconvolution and peak identification, and can also be used for quantification. To avoid quantitative issues in relation to time drift, such as variation of peak area or peak height, the detector response type has to be well defined. Due to the make-up flow and pressure regulation of make-up, the detector response (height of the peak) and peak area appeared to be dependent on experimental conditions such as inlet pressure and make-up pressure. Even if for some experimental conditions, VUV looks like mass-flow sensitive detector, it has been demonstrated that VUV is a concentration sensitive detector.

Keywords: Concentration dependent; Detector; Vacuum ultraviolet detector.

MeSH terms

  • Chromatography / instrumentation*
  • Molecular Weight
  • Pressure
  • Time Factors
  • Ultraviolet Rays*
  • Vacuum*