Hyperspectral imaging for monitoring of perfusion failure upon microvascular anastomosis in the rat hind limb

Microvasc Res. 2018 Mar:116:64-70. doi: 10.1016/j.mvr.2017.10.005. Epub 2017 Nov 6.

Abstract

Background/purpose: Objective, reliable and easy monitoring of microvascular tissue perfusion is a goal that was achieved for many years with limited success. Therefore, a new non-invasive hyperspectral camera system (TIVITA™) was tested for this purpose in an in vivo animal model.

Methods: Evaluation of tissue oxygenation during ischemia and upon reperfusion was performed in left hind limb in a rat model (n=20). Ischemia was induced by clamping and dissection of the superficial femoral artery. Reperfusion of the limb was achieved by microsurgical anastomosis of the dissected artery. Oxygenation parameters of the hind limb were assessed via TIVITA™ before and immediately after clamping and dissection of the artery, 3 and 30min after reperfusion as well as on postoperative days 1 and 2. Thereby, the non-operated hind limb served as control. As clinical parameters, the refill of the anastomosis as well as the progress of the affected leg were assessed.

Results: In 12 from 20 cases, TIVITA™ recorded a sufficient reperfusion with oxygenation parameters comparable to baseline or control condition. However, in 8 from 20 cases oxygenation was found impaired after reperfusion causing a re-assessment of the microvascular anastomosis. Thereby, technical problems like stenosis or local thrombosis were found in all cases and were surgically treated leading to an increased tissue oxygenation.

Conclusions: The TIVITA™ camera system is a valid non-invasive tool to assess tissue perfusion after microvascular anastomosis. As it safely shows problems in oxygenation, it allows the clinician a determined revision of the site in time in order to prevent prolonged ischemia.

Keywords: Hyperspectral imaging; Microvascular anastomosis; Perfusion imaging; Skin microcirculation; Tissue viability.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anastomosis, Surgical
  • Animals
  • Blood Flow Velocity
  • Femoral Artery / diagnostic imaging
  • Femoral Artery / physiopathology
  • Femoral Artery / surgery*
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / physiopathology
  • Hindlimb / blood supply*
  • Male
  • Microvessels / diagnostic imaging
  • Microvessels / physiopathology
  • Microvessels / surgery*
  • Models, Animal
  • Necrosis
  • Oxygen / blood
  • Perfusion Imaging / methods*
  • Postoperative Complications / diagnostic imaging*
  • Postoperative Complications / etiology
  • Postoperative Complications / physiopathology
  • Predictive Value of Tests
  • Rats, Sprague-Dawley
  • Regional Blood Flow
  • Reperfusion Injury / diagnostic imaging
  • Reperfusion Injury / etiology
  • Reperfusion Injury / physiopathology
  • Spectroscopy, Near-Infrared*
  • Thrombosis / diagnostic imaging
  • Thrombosis / etiology
  • Thrombosis / physiopathology
  • Time Factors
  • Tissue Survival
  • Vascular Grafting / adverse effects*

Substances

  • Oxygen