Evaluation of absolute measurement using a 4π plastic scintillator for the 4πβ-γ coincidence counting method

Appl Radiat Isot. 2018 Apr:134:302-306. doi: 10.1016/j.apradiso.2017.10.029. Epub 2017 Oct 16.

Abstract

Absolute measurement by the 4πβ-γ coincidence counting method was conducted by two photomultipliers facing across a plastic scintillator to be focused on β ray counting efficiency. The detector was held with a through-hole-type NaI(Tl) detector. The results include absolutely determined activity and its uncertainty especially about extrapolation. A comparison between the obtained and known activities showed agreement within their uncertainties.

Keywords: 4πβ-γ coincidence counting method; Absolute activity measurement; Accuracy; Co-60; Cs-134; Extrapolation; Scintillation detector; Uncertainty.