Accelerometer Detects Pump Thrombosis and Thromboembolic Events in an In vitro HVAD Circuit

ASAIO J. 2018 Sep/Oct;64(5):601-609. doi: 10.1097/MAT.0000000000000699.

Abstract

Pump thrombosis and stroke are serious complications of left ventricular assist device (LVAD) support. The aim of this study was to test the ability of an accelerometer to detect pump thrombosis and thromboembolic events (TEs) using real-time analysis of pump vibrations. An accelerometer sensor was attached to a HeartWare HVAD and tested in three in vitro experiments using different pumps for each experiment. Each experiment included thrombi injections sized 0.2-1.0 mL and control interventions: pump speed change, afterload increase, preload decrease, and saline bolus injections. A spectrogram was calculated from the accelerometer signal, and the third harmonic amplitude was used to test the sensitivity and specificity of the method. The third harmonic amplitude was compared with the pump energy consumption. The acceleration signals were of high quality. A significant change was identified in the accelerometer third harmonic during the thromboembolic interventions. The third harmonic detected thromboembolic events with higher sensitivity/specificity than LVAD energy consumption: 92%/94% vs. 72%/58%, respectively. A total of 60% of thromboembolic events led to a prolonged third harmonic amplitude change, which is indicative of thrombus mass residue on the impeller. We concluded that there is strong evidence to support the feasibility of real-time continuous LVAD monitoring for thromboembolic events and pump thrombosis using an accelerometer. Further in vivo studies are needed to confirm these promising findings.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Accelerometry / methods*
  • Animals
  • Heart-Assist Devices / adverse effects*
  • In Vitro Techniques
  • Swine
  • Thrombosis / diagnosis
  • Thrombosis / etiology*