Real-Time Spatiotemporal Control of High-Intensity Focused Ultrasound Thermal Ablation Using Echo Decorrelation Imaging in ex Vivo Bovine Liver

Ultrasound Med Biol. 2018 Jan;44(1):199-213. doi: 10.1016/j.ultrasmedbio.2017.09.007. Epub 2017 Oct 23.

Abstract

The ability to control high-intensity focused ultrasound (HIFU) thermal ablation using echo decorrelation imaging feedback was evaluated in ex vivo bovine liver. Sonications were automatically ceased when the minimum cumulative echo decorrelation within the region of interest exceeded an ablation control threshold, determined from preliminary experiments as -2.7 (log-scaled decorrelation per millisecond), corresponding to 90% specificity for local ablation prediction. Controlled HIFU thermal ablation experiments were compared with uncontrolled experiments employing two, five or nine sonication cycles. Means and standard errors of the lesion width, area and depth, as well as receiver operating characteristic curves testing ablation prediction performance, were computed for each group. Controlled trials exhibited significantly smaller average lesion area, width and treatment time than five-cycle or nine-cycle uncontrolled trials and also had significantly greater prediction capability than two-cycle uncontrolled trials. These results suggest echo decorrelation imaging is an effective approach to real-time HIFU ablation control.

Keywords: Echo decorrelation imaging; High-intensity focused ultrasound; Real-time control; Thermal ablation.

MeSH terms

  • Animals
  • Cattle
  • High-Intensity Focused Ultrasound Ablation / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Liver / surgery*