CT and MRI slice separation evaluation by LabView developed software

Z Med Phys. 2018 Feb;28(1):6-13. doi: 10.1016/j.zemedi.2017.09.009. Epub 2017 Oct 13.

Abstract

The efficient use of Computed Tomography (CT) and Magnetic Resonance Imaging (MRI) equipment necessitates establishing adequate quality-control (QC) procedures. In particular, the accuracy of slice separation, during multislices acquisition, requires scan exploration of phantoms containing test objects. To simplify such procedures, a novel phantom and a computerised LabView-based procedure have been devised, enabling determination the midpoint of full width at half maximum (FWHM) in real time while the distance from the profile midpoint of two progressive images is evaluated and measured. The results were compared with those obtained by processing the same phantom images with commercial software. To validate the proposed methodology the Fisher test was conducted on the resulting data sets. In all cases, there was no statistically significant variation between the commercial procedure and the LabView one, which can be used on any CT and MRI diagnostic devices.

Keywords: Computed Tomography; LabView application; Magnetic Resonance Imaging; Quality controls; Slice separation measurement.

Publication types

  • Validation Study

MeSH terms

  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / standards*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Software / standards*
  • Tomography, X-Ray Computed / standards*