Relevance of sample preparation for flow cytometry

Int J Lab Hematol. 2018 Apr;40(2):152-158. doi: 10.1111/ijlh.12755. Epub 2017 Oct 6.

Abstract

Introduction: Flow cytometry is a useful tool for diagnosis and minimal residual disease (MRD) study of hematological diseases. Standard sample preparation protocols are characterized by stain-lyse-wash (SLW). To prevent nonspecific bindings and achieve high sensitivity in MRD studies, lyse-wash-stain-wash (LWSW) is required. To our knowledge, no comparison between the two methods has been performed.

Methods: We compared mean fluorescence intensity (MFI), stain index, signal-to-noise ratio, and percentage of positive cells of 104 antibodies and of 13 selected antibodies tested in 10 samples simultaneously prepared with the two methods.

Results: MFI and percentages of positive cells obtained by the two methods did not show significant differences and showed a very high correlation. Stain index and signal-to-noise ratio presented higher values for kappa and lambda surface chains in LWSW samples and a trend of higher values for the other antibodies in SLW samples.

Conclusions: We suggest to use LWSW method also at diagnosis to obtain more comparable antibody intensity expressions when samples from the same patient are processed for MRD evaluation after bulk lysis. Moreover, LWSW can prevent nonspecific bindings, shows no differences in the identification and quantitation of the populations of interest, and reduces acquisition of cell debris.

Keywords: antibody binding; flow cytometry; lyse-wash-stain; sample preparation; stain-lyse-wash.

MeSH terms

  • Clinical Protocols / standards*
  • Flow Cytometry / methods*
  • Neoplasm, Residual / diagnosis
  • Sensitivity and Specificity
  • Specimen Handling / methods
  • Specimen Handling / standards