Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates

J Vis Exp. 2017 Sep 6:(127):55985. doi: 10.3791/55985.

Abstract

Amplitude-integrated EEG (aEEG) is an easily accessible technique to monitor the electrocortical activity in preterm and term infants in neonatal intensive care units (NICUs). This method was first used to monitor newborns after asphyxia, providing information about future neurological outcomes. The aEEG is also helpful to select newborns who benefit from cooling. The aEEG monitoring of preterm infants is becoming more widespread, as various studies have shown that neurodevelopmental outcome is related to early aEEG tracings. Here, we demonstrate the application of the aEEG monitoring system and present typical patterns that depend upon gestational age and pathophysiological conditions. Furthermore, we mention pitfalls in the interpretation of the aEEG, as this method requires accurate fixation and localization of the electrodes. Additionally, the raw EEG can be used to detect neonatal seizures or to identify aEEG application problems. In conclusion, aEEG is a safe and generally well-tolerated method for the bedside monitoring of neonatal cerebral function; it can even provide information about long-term outcome.

Publication types

  • Video-Audio Media

MeSH terms

  • Electroencephalography / methods*
  • Humans
  • Infant, Newborn
  • Infant, Premature
  • Intensive Care Units, Neonatal
  • Monitoring, Physiologic / methods*
  • Seizures / diagnosis*