Partial renal coverage in endovascular aneurysm repair causes unfavorable renal flow patterns in an infrarenal aneurysm model

J Vasc Surg. 2018 May;67(5):1585-1594. doi: 10.1016/j.jvs.2017.05.092. Epub 2017 Oct 12.

Abstract

Objective: To achieve an optimal sealing zone during endovascular aneurysm repair, the intended positioning of the proximal end of the endograft fabric should be as close as possible to the most caudal edge of the renal arteries. Some endografts exhibit a small offset between the radiopaque markers and the proximal fabric edge. Unintended partial renal artery coverage may thus occur. This study investigated the consequences of partial coverage on renal flow patterns and wall shear stress (WSS).

Methods: In vitro models of an abdominal aortic aneurysm were used to visualize pulsatile flow using two-dimensional particle image velocimetry under physiologic resting conditions. One model served as control and two models were stented with an Endurant endograft (Medtronic Inc, Minneapolis, Minn), one without and one with partial renal artery coverage with 1.3 mm of stent fabric extending beyond the marker (16% area coverage). The magnitude and oscillation of WSS, relative residence time, and backflow in the renal artery were analyzed.

Results: In both stented models, a region along the caudal renal artery wall presented with low and oscillating WSS, not present in the control model. A region with very low WSS (<0.1 Pa) was present in the model with partial coverage over a length of 7 mm compared with a length of 2 mm in the model without renal coverage. Average renal backflow area percentage in the renal artery incrementally increased from control (0.9%) to the stented model without (6.4%) and with renal coverage (18.8%).

Conclusions: In this flow model, partial renal coverage after endovascular aneurysm repair causes low and marked oscillations in WSS, potentially promoting atherosclerosis and subsequent renal artery stenosis. Awareness of the device-dependent offset between the fabric edge and the radiopaque markers is therefore important in endovascular practice.

Publication types

  • Research Support, Non-U.S. Gov't
  • Video-Audio Media

MeSH terms

  • Aorta, Abdominal / diagnostic imaging
  • Aorta, Abdominal / physiopathology
  • Aorta, Abdominal / surgery*
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / physiopathology
  • Aortic Aneurysm, Abdominal / surgery*
  • Blood Flow Velocity
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation / adverse effects*
  • Blood Vessel Prosthesis Implantation / instrumentation
  • Computed Tomography Angiography
  • Endovascular Procedures / adverse effects*
  • Endovascular Procedures / instrumentation
  • Humans
  • Models, Anatomic*
  • Models, Cardiovascular*
  • Prosthesis Design
  • Renal Artery / diagnostic imaging
  • Renal Artery / physiopathology
  • Renal Artery / surgery*
  • Renal Artery Obstruction / etiology
  • Renal Artery Obstruction / physiopathology
  • Renal Circulation*
  • Risk Factors
  • Stents
  • Stress, Mechanical
  • Time Factors