Objective: The aim of this study was to evaluate the accuracy of cone beam computed tomography (CBCT) for evaluation of the bone-implant interface in comparison with periapical radiography.
Study design: Titanium implants were inserted in 74 bovine rib blocks in intimate contact with bone walls and with a gap of 0.125 mm (simulating failure in the osseointegration process). Periapical radiographs were taken with conventional film, and CBCT scans were acquired with i-CAT (0.2 mm and 0.125 mm voxel) and Kodak (0.2 mm and 0.076 mm voxel) units. Three examiners evaluated the images using a 5-point scale. Diagnostic accuracy was analyzed through sensitivity, specificity, and the area under the receiver operating characteristic (ROC) curve (AUC) with 95% confidence intervals (CIs). Intra- and interexaminer agreements were analyzed through Kendall's concordance test.
Results: Intra- and interexaminer agreements showed satisfactory results. The greatest accuracy was observed with conventional radiography (AUC = 0.963; CI 95% = 0.891-0.993). I-CAT 0.125-mm images showed good accuracy (AUC = 0.885; CI 95% = 0.790-0.947), with no significant difference compared with conventional radiography. Kodak images had high specificity and low sensitivity, presenting more false-negative results.
Conclusions: Conventional radiography showed the highest accuracy for assessment of the bone-implant interface. However, CBCT (i-CAT; 0.125-mm voxel), if available or if performed for preoperative assessment of another implant site, may provide similar accuracy.