Operational impact of using a vanadate oxidase method for direct bilirubin measurements at an academic medical center clinical laboratory

Pract Lab Med. 2017 May 17:8:77-85. doi: 10.1016/j.plabm.2017.05.004. eCollection 2017 Aug.

Abstract

Objectives: The aim of this study was to compare the operational impact of using vanadate oxidase versus diazo direct bilirubin assays for an academic medical center patient population.

Design and methods: Retrospective study was done over an approximately 3.5 year period. The main automated chemistry instrumentation was a Roche Diagnostics cobas 8000 line. The Roche Direct Bilirubin assay was compared to Diazyme Laboratories Direct Bilirubin Assay and Randox Laboratories Direct Bilirubin assay using manufacturer's guidelines for hemolysis index, lipemia index, and analytical measurement range (AMR).

Results: Retrospective data was analyzed for 47,333 serum/plasma specimens that had clinical orders for direct bilirubin. A total of 5943 specimens (12.6%) exceeded the hemolysis index limit for the Roche method compared to only 0.2% and 0.05% of specimens for the Diazyme and Randox methods, respectively. The impact was particularly large on patients less than 2 years old, for which 51.3% of specimens exceeded the hemolysis index for the Roche method. A total of 1671 specimens (3.5%) exceeded the lipemia index limit for the Roche method compared to less than 0.1% for the Randox method. Lastly, 988 (2.1%) of specimens had direct bilirubin concentrations exceeding the upper AMR limit of 10 mg/dL [171 µmol/L] for the Roche assay compared to less than 1% of specimens for the vanadate oxidase methods.

Conclusions: Vanadate oxidase direct bilirubin methods offer advantages over diazo methods in terms of less interference by hemolysis and lipemia, as well as wider AMR. The advantages are particularly evident for neonatal and infant populations.

Keywords: Bilirubin; Clinical chemistry tests; Hemolysis; Hyperlipidemias; Jaundice; Photometry.