Comparison of hemodialysis with medium cut-off dialyzer and on-line hemodiafiltration on the removal of small and middle-sized molecules

Clin Nephrol. 2018 Jan;89 (2018)(1):50-56. doi: 10.5414/CN109133
.

Abstract

Background: Recent data suggest that the use of medium cut-off (MCO) dialyzers in hemodialysis (HD) promotes greater clearance and reduction ratio (RR) for myoglobin and other large-sized molecules than on-line hemodiafiltration (ol-HDF), but its effects on β2-microglobulin are not clear. We compared RR and clearances of small and middle-sized molecules between high-flux ol-HDF and MCO (Theranova) dialyzer in HD (MCO-HD) as well as nutritional parameters.

Materials and methods: We retrospectively analyzed 10 patients treated first with ol-HDF who were thereafter switched to MCO-HD over a 1-year period. Three dialysis sessions in each 6-month period were examined. We calculated RR and clearance of small and middle-sized molecules.

Results: There was no significant difference between ol-HDF and MCO-HD for median serum albumin and prealbumin level, mean KT/V, mean urea and creatinine RR, mean β2-microglobulin (81 ± 5 vs. 81 ± 6%, p = 0.72) and myoglobin (60 ± 9% vs. 61 ± 7%, p = 0.59), RR or clearances.

Conclusion: The use of MCO (Theranova) dialyzer in HD produces similar removal of urea, creatinine, β2-microglobulin and myoglobin as does ol-HDF, with good tolerance profile and without modification of nutritional status. .

MeSH terms

  • Creatinine / analysis
  • Hemodiafiltration* / instrumentation
  • Hemodiafiltration* / methods
  • Hemodiafiltration* / statistics & numerical data
  • Humans
  • Renal Dialysis* / instrumentation
  • Renal Dialysis* / methods
  • Renal Dialysis* / statistics & numerical data
  • Retrospective Studies
  • Urea / analysis
  • beta 2-Microglobulin / analysis

Substances

  • beta 2-Microglobulin
  • Urea
  • Creatinine