Objective: To determine the comparison of its clinical utility and safety profile for visual evoked potential (VEP) monitoring during prone spine surgeries under total intravenous anesthesia (TIVA) versus balanced general anesthesia using the SightSaver™ visual stimulator.
Methods: The protocol was designed asa pilot, single center, prospective, randomized, and double-arm study. Subjects were randomized to receive either TIVA or balanced general anesthesia. Following induction and intubation, 8 electrodes were placed subcutaneously to collect VEP recordings. The SightSaver™ visual stimulator was placed on the subject's scalp before prone positioning. VEP waveforms were recorded every 30min and assessed by a neurophysiologist throughout the length of surgery.
Results: A total of 19 subjects were evaluated and VEP waveforms were successfully collected. TIVA group showed higher amplitude and lower latency than balanced anesthesia.
Conclusions: Our data suggested that TIVA is associated with higher VEP amplitude and shorter latencies than balanced general anesthesia; therefore, TIVA could be the most efficient anesthesia regimen for VEP monitoring.
Significance: The findings help to better understand the effect of different anesthesia regimens on intra-operative VEP monitoring.
Keywords: Balanced general anesthesia; Prone spinal surgery; SightSaver™; Total intravenous anesthesia; Visual evoked potential.
Copyright © 2017 International Federation of Clinical Neurophysiology. Published by Elsevier B.V. All rights reserved.