Evaluation of the radon interference on the performance of the portable monitoring air pump for radioactive aerosols (MARE)

Appl Radiat Isot. 2018 Apr:134:439-445. doi: 10.1016/j.apradiso.2017.07.039. Epub 2017 Jul 27.

Abstract

A compact portable aerosol sampling and measurement device was developed at Jožef Stefan Institute (JSI). A CeBr3 scintillation detector is positioned centrally within a concertinaed filter assembly. It provides continuous and via network communications on-line monitoring of low levels of airborne radioactive particulates. The evaluation of the response of the device to the natural background at controlled conditions with elevated radon concentrations, performed at the National Institute of Ionizing Radiation Metrology of ENEA, is presented.

Keywords: Aerosol sampling device; CeBr(3) scintillation detector; Early warning network; High volume air sampler; MetroERM; Preparedness; Radiological emergency; Radon and thoron background; Radon progeny; Real time measurements; Walk-in radon chamber.