The interplay of electrode- and bio-materials in a redox-cycling-based clozapine sensor

Electrochem commun. 2017 Jun:79:33-36. doi: 10.1016/j.elecom.2017.04.009. Epub 2017 Apr 13.

Abstract

We investigate gold, TiN, and platinum in combination with a chitosan-catechol-based redox-cycling system (RCS) for electrochemical detection of the antipsychotic clozapine. We have previously demonstrated the RCS for detection of clozapine in serum, but challenges remain regarding low signal-to-noise ratios. This can be mitigated by selection of electrode materials with beneficial surface morphologies and/or compositions. We employ cyclic voltammetry to assess the redox current generated by clozapine, and differentiate solely surface-area-based effects from clozapine-specific ones using a standard redox couple. We find that nano- and microstructured platinum greatly amplifies the clozapine signal compared to gold (up to 1490-fold for platinum black). However, the material performs poorly in the presence of chloride ions, and RCS modification provides no further amplification. The RCS combined with atomic-layer-deposited (ALD) TiN, on the other hand, increases the signal by 7.54 times, versus 2.86 times for RCS on gold, with a 9.2-fold lower variability, indicating that the homogenous and chemically inert properties of ALD-TiN may make it an ideal electrode material.

Keywords: clozapine; platinum black; redox cycling; therapeutic drug monitoring; titanium nitride.