Results of the CCRI(II)-S12.H-3 supplementary comparison: Comparison of methods for the calculation of the activity and standard uncertainty of a tritiated-water source measured using the LSC-TDCR method

Appl Radiat Isot. 2018 Apr:134:257-262. doi: 10.1016/j.apradiso.2017.07.007. Epub 2017 Jul 6.

Abstract

A comparison of calculations of the activity of a 3H2O liquid scintillation source using the same experimental data set collected at the LNE-LNHB with a triple-to-double coincidence ratio (TDCR) counter was completed. A total of 17 laboratories calculated the activity and standard uncertainty of the LS source using the files with experimental data provided by the LNE-LNHB. The results as well as relevant information on the computation techniques are presented and analysed in this paper. All results are compatible, even if there is a significant dispersion between the reported uncertainties. An output of this comparison is the estimation of the dispersion of TDCR measurement results when measurement conditions are well defined.

Keywords: Activity comparison; LSC; TDCR; Tritium.