Multistage stereotactic radiosurgery for large cerebral arteriovenous malformations using the Gamma Knife platform

Med Phys. 2017 Oct;44(10):5010-5019. doi: 10.1002/mp.12455. Epub 2017 Aug 12.

Abstract

Purpose: Radiosurgery is an established technique to treat cerebral arteriovenous malformations (AVMs). Obliteration of larger AVMs (> 10-15 cm3 or diameter > 3 cm) in a single session is challenging with current radiosurgery platforms due to toxicity. We present a novel technique of multistage stereotactic radiosurgery (SRS) for large intracranial arteriovenous malformations (AVM) using the Gamma Knife system.

Materials/methods: Eighteen patients with large (> 10-15 cm3 or diameter > 3 cm) AVMs, which were previously treated using a staged SRS technique on the Cyberknife platform, were retrospectively selected for this study. The AVMs were contoured and divided into 3-8 subtargets to be treated sequentially in a staged approach at half to 4 week intervals. The prescription dose ranged from 15 Gy to 20 Gy, depending on the subtarget number, volume, and location. Gamma Knife plans using multiple collimator settings were generated and optimized. The coordinates of each shot from the initial plan covering the total AVM target were extracted based on their relative positions within the frame system. The shots were regrouped based on their location with respect to the subtarget contours to generate subplans for each stage. The delivery time of each shot for a subtarget was decay corrected with 60 Co for staging the treatment course to generate the same dose distribution as that planned for the total AVM target. Conformality indices and dose-volume analysis were performed to evaluate treatment plans.

Results: With the shot redistribution technique, the composite dose for the multistaged treatment of multiple subtargets is equivalent to the initial plan for total AVM target. Gamma Knife plans resulted in an average PTV coverage of 96.3 ± 0.9% and a PITV of 1.23 ± 0.1. The resulting Conformality indices, V12Gy and R50 dose spillage values were 0.76 ± 0.05, 3.4 ± 1.8, and 3.1 ± 0.5 respectively.

Conclusion: The Gamma Knife system can deliver a multistaged conformal dose to treat large AVMs when correcting for translational setup errors of each shot at each staged treatment.

Keywords: Gamma Knife; arteriovenous malformation; stereotactic radiosurgery.

MeSH terms

  • Adolescent
  • Adult
  • Arteriovenous Fistula / radiotherapy*
  • Child
  • Female
  • Humans
  • Intracranial Arteriovenous Malformations / radiotherapy*
  • Male
  • Middle Aged
  • Radiosurgery / instrumentation*
  • Radiotherapy Planning, Computer-Assisted
  • Retrospective Studies
  • Treatment Outcome
  • Young Adult