Validity and reliability of the myTemp ingestible temperature capsule

J Sci Med Sport. 2018 Mar;21(3):322-326. doi: 10.1016/j.jsams.2017.06.006. Epub 2017 Jun 8.

Abstract

Objectives: An accurate and non-invasive measurement of core body temperature (Tc) is of great importance to quantify exercise-induced increases in Tc in athletes or to assess changes in Tc in patient populations. The use of ingestible gastrointestinal telemetric temperature capsules is widely accepted as a surrogate marker for Tc, but widespread implementation is lacking due to the high costs of these disposable capsules. A new and cheaper temperature capsule system (i.e. myTemp) was recently introduced. The aim of present study is to determine the validity and test-retest reliability of the myTemp system.

Design: Ex-vivo experimental study.

Methods: Fifteen ingestible temperature capsules (myTemp, Nijmegen, Netherlands) were tested in a highly temperature controlled water bath, in which the water temperature gradually increased from 34°C to 44°C. The study protocol was performed twice for each temperature capsule.

Results: Mean difference between myTemp temperature and water bath temperature was -0.001±0.005°C (Limit of Agreement (LOA): ±0.011°C) during Trial 1 (p=0.11) and -0.001±0.006°C (LOA: ±0.012°C) during Trial 2 (p=0.039). Furthermore, an Intraclass Correlation Coefficient (ICC) of 1.00 was found for both trials. A systematic difference between Trials 1 and 2 of 0.004±0.008°C (LOA: ±0.015°C) was found (p<0.001), whereas the ICC between both trials was 1.00 and the standard error of measurement was 0.005°C.

Conclusions: Although we found a systematic bias for the sensitivity (-0.001°C) and reliability (0.004°C), these values can be considered insignificant from a physiological and clinical perspective. Thus, the myTemp ingestible temperature capsule is a valid technique to measure (water) temperature under controlled circumstances.

Keywords: Core body temperature; Intestinal temperature; Thermometer; Thermoregulation.

Publication types

  • Validation Study

MeSH terms

  • Analysis of Variance
  • Body Temperature*
  • Capsules
  • Exercise / physiology*
  • Hot Temperature*
  • Humans
  • Reproducibility of Results
  • Telemetry
  • Water