Reconstruction of AAOS type III and IV acetabular defects with the Ganz reinforcement ring: high failure in pelvic discontinuity

Arch Orthop Trauma Surg. 2017 Aug;137(8):1139-1148. doi: 10.1007/s00402-017-2731-x. Epub 2017 Jun 16.

Abstract

Background: Large acetabular defects and pelvic discontinuity represent complex problems in revision total hip arthroplasty. This study aimed to investigate whether reconstruction with the Ganz reinforcement ring would provide durable function in large acetabular defects.

Patients and methods: 46 hips (45 patients, 19 male, 26 female, mean age 68 years) with AAOS type III and IV defects undergoing acetabular revision with the Ganz reinforcement ring were evaluated at a mean follow-up of 74 months (24-161 months). Fourteen patients died during follow-up. All surviving patients were available for clinical assessment and radiographic studies. Radiographs were evaluated for bone healing and component loosening. A Cox-regression model was performed to identify factors influencing survival of the Ganz-ring.

Results: In the group of AAOS III defects, 3 of 26 acetabular reconstructions failed, all due to aseptic loosening. In pelvic discontinuity (AAOS IV), 9 of 20 hips failed due to aseptic loosening (n = 4), deep infection (n = 3), and non-union of the pelvic ring (n = 2). With acetabular revision for any reason as an endpoint, the estimated Kaplan-Meier 5-year survival was 86% in type III defects and 57% in type IV defects, respectively. The presence of pelvic discontinuity was identified as the only independent predictive factor for failure of the Ganz ring acetabular reconstruction (AAOS III vs. IV, Hazard ratio: 0.217, 95%, Confidence interval: 0.054-0.880, p = 0.032).

Conclusion: The Ganz reinforcement ring remains a favorable implant for combined segmental and cavitary defects. However, defects with pelvic discontinuity demonstrate high failure rates. The indications should therefore be narrowed to acetabular defects not associated with pelvic discontinuity.

Keywords: Acetabular defect; Acetabular reconstruction; Bone loss; Cage; Ganz reinforcement ring; Hip revision; Pelvic discontinuity; Trabecular metal.

MeSH terms

  • Acetabulum* / injuries
  • Acetabulum* / surgery
  • Aged
  • Arthroplasty, Replacement, Hip* / adverse effects
  • Arthroplasty, Replacement, Hip* / instrumentation
  • Arthroplasty, Replacement, Hip* / methods
  • Arthroplasty, Replacement, Hip* / statistics & numerical data
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Reoperation* / adverse effects
  • Reoperation* / instrumentation
  • Reoperation* / methods
  • Reoperation* / statistics & numerical data