Clinical screening of paraquat in plasma samples using capillary electrophoresis with contactless conductivity detection: Towards rapid diagnosis and therapeutic treatment of acute paraquat poisoning in Vietnam

J Chromatogr B Analyt Technol Biomed Life Sci. 2017 Aug 15:1060:111-117. doi: 10.1016/j.jchromb.2017.06.010. Epub 2017 Jun 12.

Abstract

The employment of a purpose-made capillary electrophoresis (CE) instrument with capacitively coupled contactless conductivity detection (C4D) as a simple and cost-effective solution for clinical screening of paraquat in plasma samples for early-stage diagnosis of acute herbicide poisoning is reported. Paraquat was determined using an electrolyte composed of 10mM histidine adjusted to pH 4 with acetic acid. A detection limit of 0.5mg/L was achieved. Good agreement between results from CE-C4D and the confirmation method (HPLC-UV) was obtained, with relative errors for the two pairs of data better than 20% for 31 samples taken from paraquat-intoxicated patients. The results were used by medical doctors for identification and prognosis of acute paraquat poisoning cases. The objective of the work is the deployment of the developed approach in rural areas in Vietnam as a low-cost solution to reduce the mortality rate due to accidental or suicidal ingestion of paraquat.

Keywords: Capacitively coupled contactless conductivity detection (C(4)D); Capillary electrophoresis (CE); Paraquat; Plasma samples; Purpose-made instrument; Vietnam.

MeSH terms

  • Electric Conductivity
  • Electrophoresis, Capillary / methods*
  • Hemoperfusion
  • Humans
  • Limit of Detection
  • Paraquat / blood*
  • Paraquat / poisoning
  • Poisoning* / blood
  • Poisoning* / diagnosis
  • Poisoning* / therapy
  • Reproducibility of Results
  • Vietnam

Substances

  • Paraquat